52k+ listingsShips to NLResponse within one hour+31 73 711 4717info@questpair.com

Bio-Rad Gene Pulser Transfection Apparatus with 27 Cuvettes P/N 165-2089

From€895.77excl. VAT

In stock

Can't find what you're looking for?

Application
Mammalian Cells, Plant Protoplasts
Max Voltage
2500 V
Sample Format
Cuvette
All 7 specifications
  • Ships to NL
  • Response within 1 hour
  • Warranty available. Buy without worry
  • Support from a scientist

Specifications 7

Weight
13.2
Dimensions
26.7 × 30.5 × 45.7
Brands
Bio-Rad
New/Used
Used
Application
Mammalian Cells, Plant Protoplasts
Max Voltage
2500 V
Sample Format
Cuvette

Product description

Bio-Rad Gene Pulser electroporator with 27 new P/N 165-2089 cuvettes for mammalian cell and plant protoplast electroporation.

The Bio-Rad Gene Pulser electroporator is a capacitor discharge pulse generator for cell electroporation. It is recommended for mammalian cells, plant protoplasts and other eukaryotic cells.

The unit includes 27 new and unused cuvettes, P/N 165-2089. Rear connectors support an optional Capacitance Extender. The specific 0.2 cm and 0.4 cm cuvettes referenced for field strength are not included.

The unit is used. The included cuvettes are new and unused.

The Bio-Rad Gene Pulser is a capacitor discharge electroporator for controlled exponential pulses in cell electroporation workflows.

  • Pulse voltage up to 2,500 volts
  • Current above 125 amperes
  • Field strength up to 12,500 V/cm with a 0.2 cm cuvette
  • Field strength up to 6,250 V/cm with a 0.4 cm cuvette
  • Digital readout for voltage and capacitance settings
  • Measurement and display of actual voltage after each pulse
  • Measurement and display of the RC time constant after each pulse
  • Internal capacitance values of 0.25, 1.0, 3.0 and 25.0 µF
  • Rear connectors for an optional Capacitance Extender supporting 125, 250, 500 and 960 µF at up to 450 volts
  • Suitable for mammalian cell and plant protoplast electroporation
  • Includes 27 new and unused cuvettes, P/N 165-2089

The unit is used. The 0.2 cm and 0.4 cm cuvettes referenced for field strength are not included.

What these specifications mean

The pulse voltage reaches 2,500 volts and the field strength reaches 12,500 V/cm with a 0.2 cm cuvette or 6,250 V/cm with a 0.4 cm cuvette. The smaller cuvette gap therefore provides twice the maximum field strength of the larger gap.

The internal capacitors provide 0.25, 1.0, 3.0 and 25.0 µF settings. The optional Capacitance Extender expands the capacitance range to 125, 250, 500 and 960 µF, with a maximum voltage of 450 volts.

Why Questpair exists

Lab equipment markets are fragmented. Buyers spend weeks calling distributors, comparing quotes, and verifying supplier reliability. Sellers struggle to reach global research labs.

We built Questpair to fix this. Every supplier on the platform is verified before listing. Every quote you receive comes from a real, vetted seller. AI prepares your case in minutes, then a real specialist takes over.

We're an independent team based in The Netherlands, working with research labs and procurement teams across Europe and worldwide.

Frequently asked questions

How long does laboratory equipment typically last?

With a service contract, ten to fifteen years is normal for most benchtop instruments. Detectors and pumps are the parts that wear first.

Can I get service for older equipment?

Often, yes. Verified suppliers frequently work with service partners converting the instrument. If applicable, we can list servicing options alongside the offer.

Where can I buy parts?

Ask a scientist through the request button. Parts are often available even when the instrument itself is discontinued.

What is the typical lead time for refurbished equipment?

Two to six weeks, depending on refurbishment scope and whether calibration is included.

Does laboratory equipment ship internationally?

We ship to 38 countries. Crating and customs handling are quoted with the instrument.

Reply in one hour

Can't find the exact configuration?

Tell us what you need and a scientist finds it for you, usually within the hour.

Last updated 28 November 2026