Max Processing Volume
20 L
Max Viscosity
10000 mPa·s
Max Temperature
350 °C
All 8 specifications
  • Ships to NL
  • Response within 1 hour
  • Warranty available. Buy without worry
  • Support from a scientist

Specifications 8

Weight
5.2
Dimensions
19.1 × 8.9 × 30.6
Brands
IKA
New/Used
Unused
Stirrer Type
Overhead Stirrer
Max Processing Volume
20 L
Max Viscosity
10000 mPa·s
Max Temperature
350 °C

Product description

IKA EUROSTAR 20 high speed control overhead stirrer with R 1405 propeller, H 67.61 temperature sensor and 6000 rpm speed.

The IKA EUROSTAR 20 high speed control overhead stirrer is designed for sample mixing in biotechnology and life sciences laboratories. It provides controlled stirring for liquid processing and formulation work.

The unit includes an R 1405 Propeller and an H 67.61 Temperature Sensor. The propeller provides the stirring element, while the temperature sensor adds temperature monitoring during operation. A tripod is not included.

The unit is unused. The H 67.61 temperature sensor has a maximum temperature of 290 °C.

The IKA EUROSTAR 20 high speed control is an overhead stirrer for laboratory sample mixing. The unused unit includes an R 1405 Propeller and an H 67.61 Temperature Sensor.

Performance and control

  • Maximum stirring quantity per position with water: 20 L
  • Speed range: 0 to 150 to 6000 rpm
  • Speed setting accuracy: ±1 rpm
  • Maximum viscosity: 10,000 mPa·s
  • Maximum torque at the stirring shaft: 40 Ncm
  • Maximum torque at the stirring shaft from 150 to 600 rpm: 40 Ncm
  • Maximum torque at the stirring shaft from 600 to 6000 rpm: 40 to 20 Ncm
  • Nominal torque: 0.2 Nm
  • Torque measurement deviation: ±1 Ncm
  • Speed measurement deviation above 300 rpm: ±1%
  • Speed measurement deviation below 300 rpm: ±3 rpm
  • Time setting range: 1 to 6000 min

Temperature measurement

  • External temperature sensor connection: PT1000
  • Temperature measurement range: -10 to 350 °C
  • Temperature measurement resolution: 0.1 K
  • Temperature measurement accuracy: ±0.5 + PT1000 tolerance, DIN EN 60751 Class A, K
  • Temperature sensor limit deviation: less than or equal to ±(0.15 + 0.002 x ITI) K
  • H 67.61 temperature sensor maximum temperature: 290 °C

Included components and construction

  • R 1405 Propeller
  • R 1405 working range: 0.25 to 30 L
  • H 67.61 Temperature Sensor
  • Housing material: aluminium cast coating and thermoplastic polymer
  • Dimensions, W x H x D: 89 x 306 x 191 mm
  • Weight: 5.2 kg
  • Tripod not included

What these specifications mean

The 10,000 mPa·s maximum viscosity supports mixing of more viscous samples than applications limited to low-viscosity liquids. The 20 L maximum quantity per stirring position places the unit within small batch and laboratory-scale mixing work.

The 6000 rpm upper speed supports high-speed agitation. The included H 67.61 sensor is rated to 290 °C, so the accessory operates below the stirrer's broader temperature measurement range of -10 to 350 °C.

Why Questpair exists

Lab equipment markets are fragmented. Buyers spend weeks calling distributors, comparing quotes, and verifying supplier reliability. Sellers struggle to reach global research labs.

We built Questpair to fix this. Every supplier on the platform is verified before listing. Every quote you receive comes from a real, vetted seller. AI prepares your case in minutes, then a real specialist takes over.

We're an independent team based in The Netherlands, working with research labs and procurement teams across Europe and worldwide.

Frequently asked questions

How long does laboratory equipment typically last?

With a service contract, ten to fifteen years is normal for most benchtop instruments. Detectors and pumps are the parts that wear first.

Can I get service for older equipment?

Often, yes. Verified suppliers frequently work with service partners converting the instrument. If applicable, we can list servicing options alongside the offer.

Where can I buy parts?

Ask a scientist through the request button. Parts are often available even when the instrument itself is discontinued.

What is the typical lead time for refurbished equipment?

Two to six weeks, depending on refurbishment scope and whether calibration is included.

Does laboratory equipment ship internationally?

We ship to 38 countries. Crating and customs handling are quoted with the instrument.

Reply in one hour

Can't find the exact configuration?

Tell us what you need and a scientist finds it for you, usually within the hour.

Last updated 24 November 2026