Applied Biosystems 7300 Real Time PCR System with PC & Software Lab
This Applied Biosystems 7300 Real Time PCR System was removed from a university lab where it was sur (more)plus to requirements. The system is comprised of; 1 x Applied Biosystems 7300 Real-Time PCR Unit. 1 x Dell Optiplex 380 with Monitor, keyboard, mouse, USB cable & monitor cable.Running Applied Biosystems Sequence Detection Software (SDS)Version 1.2.3.The Dell has a Fresh install of Windows XP. The 7300 has had a function test run: USB - PassCCD - Pass (version 2.19)Filter Wheel - Pass (FW version G2.10)Shutter - Pass (FW version G2.10)Lamp - Pass (FW version G2.10)Thermal Cycler - Pass (G2.10 Firmware is up to date) The lamp status is Good.It has 1444 Usage hours.It was last replaced 08/01/2008. The unit will require the ROI calibration.