51k+ listingsShips to NLResponse within one hour+31 73 711 4717info@questpair.com
Sold · specialists can find another

Agilent 1260 Infinity I HPLC G1329B Autosampler

UsedAgilentHPLC autosamplersUnited States

Sold

This item is sold

Our specialists can find you another one, the same model or a close equivalent, and send you a quote.

Pressure
Operating range 0 - 60 MPa (0 - 600 bar, 0 - 8850 psi)
GLP Features
Early maintenance feedback (EMF), electronic records of maintenance and errors
Communications
Controller-area network (CAN), RS232C, APG-remote standard, optional four external contact closures and BCD vial number output
Safety Features
Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display
Injection Range
0.1 – 100 µl in 0.1 µl increments (recommended 1 µl increments) Up to 1500 µl with multiple draw (hardware modification required)
Availability
Sold
All 18 specifications
  • Ships to NL
  • Response within 1 hour
  • Warranty available. Buy without worry
  • Support from a scientist

Specifications 18

Brands
Agilent
New/Used
Used
Pressure
Operating range 0 - 60 MPa (0 - 600 bar, 0 - 8850 psi)
GLP Features
Early maintenance feedback (EMF), electronic records of maintenance and errors
Communications
Controller-area network (CAN), RS232C, APG-remote standard, optional four external contact closures and BCD vial number output
Safety Features
Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display
Injection Range
0.1 – 100 µl in 0.1 µl increments (recommended 1 µl increments) Up to 1500 µl with multiple draw (hardware modification required)
Replicate Injections
1 – 99 from one vial
Precision
Typically < 0.25 % RSD of peak areas from 5 to 100 µl Typically < 1 % RSD of peak areas from 1 to 5 µl
Minimum Sample Volume
1 µl from 5 µl sample in 100 µl microvial, or 1 µl from 10 µl sample in 300 µl microvial
Carryover
Typically < 0.1 %, < 0.05 % with external needle cleaning
Sample Viscosity Range
0.2 – 50 cp
Metering Devices
Metering pump in high pressure flow path
Sample Capacity
100 × 2 ml vials in 1 tray 40 × 2 ml vials in ½ tray 15 × 6 ml vials in ½ tray (Agilent vials only)
Injection Cycle Time
50 seconds for draw speed 200 µl/min, ejection speed 200 µl/min, injection volume 5 µl
Ambient Operating Temperature
0 to 55°C
Dimensions (H x W x D)
8 x 13.5 x 17 in (20 x 34.5 x 43.5 cm)
Weight
32 lb. (14.2 kg)

Product description

HPLC autosampler with 0.1–100 µl injection range, X-Z-Theta transport, 60 MPa pressure limit.

Agilent 1260 Infinity I HPLC G1329B Autosampler Specifications and Features:

  • Injection Range: 0.1 – 100 µl
  • Injection Precision: 0.25% RSD of peak areas (5 to 100 µl)
  • Pressure Range: 0 – 60 MPa (600 bar)
  • Sample Capacity: Up to 100 samples
  • Transportation Mechanism: X-Z-Theta movement optimizing vial pick-up and return via a gripper arm
  • Flow-through Design: Continuous flow-through ensures sample loop, needle interior, seat capillary, and injection valve mainpass remain in flow line, minimizing carryover
  • Leak Handling: Leak detection with safe leak handling and error display
  • Maintenance: Early maintenance feedback (EMF) with electronic records of maintenance and errors
  • Control Compatibility: Operated via Agilent Instant Pilot or other Agilent control software
  • Pressure Operation: Can be operated up to 600 bar
  • Optional Pairing: Compatible with Agilent 1290 Infinity Thermostat (G1330B) for temperature control applications
  • Monitoring: Optical sensors and optical encoders monitor transport movement ensuring correct operation

This autosampler integrates seamlessly with Agilent 1200 Infinity Series, 1200 Series, 1100 Series LC systems, or other LC systems with remote control capabilities. Fully factory tested to meet performance specifications prior to shipment.

Why Questpair exists

Lab equipment markets are fragmented. Buyers spend weeks calling distributors, comparing quotes, and verifying supplier reliability. Sellers struggle to reach global research labs.

We built Questpair to fix this. Every supplier on the platform is verified before listing. Every quote you receive comes from a real, vetted seller. AI prepares your case in minutes, then a real specialist takes over.

We're an independent team based in The Netherlands, working with research labs and procurement teams across Europe and worldwide.

Frequently asked questions

How long does laboratory equipment typically last?

With a service contract, ten to fifteen years is normal for most benchtop instruments. Detectors and pumps are the parts that wear first.

Can I get service for older equipment?

Often, yes. Verified suppliers frequently work with service partners converting the instrument. If applicable, we can list servicing options alongside the offer.

Where can I buy parts?

Ask a scientist through the request button. Parts are often available even when the instrument itself is discontinued.

What is the typical lead time for refurbished equipment?

Two to six weeks, depending on refurbishment scope and whether calibration is included.

Does laboratory equipment ship internationally?

We ship to 38 countries. Crating and customs handling are quoted with the instrument.

Reply in one hour

Can't find the exact configuration?

Tell us what you need and a scientist finds it for you, usually within the hour.

Last updated 24 November 2026