52k+ listingsShips to NLResponse within one hour+31 73 711 4717info@questpair.com
Sold · specialists can find another

Agilent 1200 HPLC G1329A Variable Volume Autosampler

UsedAgilentHPLC autosamplersUnited States

Sold

This item is sold

Our specialists can find you another one, the same model or a close equivalent, and send you a quote.

GLP Features
Early maintenance feedback (EMF), electronic records of maintenance and errors
Communications
Controller-area network (CAN). GPIB (IEEE-448), RS232C, APG-remote standard, optional four external contact closures and BCD vial number output
Safety Features
Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display
Injection Range
0.1 – 100 µl in 0.1 µl increments Up to 1500 µl with multiple draw (hardware modification required)
Replicate Injections
1 – 99 from one vial
Availability
Sold
All 14 specifications
  • Ships to NL
  • Response within 1 hour
  • Warranty available. Buy without worry
  • Support from a scientist

Specifications 14

Brands
Agilent
New/Used
Used
GLP Features
Early maintenance feedback (EMF), electronic records of maintenance and errors
Communications
Controller-area network (CAN). GPIB (IEEE-448), RS232C, APG-remote standard, optional four external contact closures and BCD vial number output
Safety Features
Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display
Injection Range
0.1 – 100 µl in 0.1 µl increments Up to 1500 µl with multiple draw (hardware modification required)
Replicate Injections
1 – 99 from one vial
Precision
Typically < 0.5 % RSD of peak areas from 5 – 100 µl Typically < 1 % RSD 1 – 5 µl variable volume
Minimum Sample Volume
1 µl from 5 µl sample in 100 µl microvial, or 1 µl from 10 µl sample in 300 µl microvial
Carryover
Typically < 0.1 %, < 0.05 % with external needle cleaning
Sample Viscosity Range
0.2 – 50 cp
Replicate Injections Per Vial
1 – 99
Sample Capacity
100 × 2-ml vials in 1 tray 40 × 2-ml vials in ½ tray 15 × 6-ml vials in ½ tray (Agilent vials only)
Injection Cycle Time
Typically 50 s depending on draw speed and injection volume

Product description

Variable volume autosampler for HPLC with 0.1–100 µl injection range and 100-sample capacity

Agilent 1200 HPLC G1329A Autosampler

  • Injection Range: 0.1 – 100 µl in 0.1 µl increments
  • Precision:
  • Typically <0.5% RSD of peak areas for 5 – 100 µl
  • Typically <1% RSD for 1 – 5 µl variable volume
  • Sample Capacity:
  • 100 × 2 ml vials (1 tray)
  • 40 × 2 ml vials (½ tray)
  • 15 × 6 ml vials (½ tray, Agilent vials only)
  • Temperature Control: Programmable, 4°C to 40°C with -1°C to +5°C accuracy
  • Features:
  • GLP compliant with early maintenance feedback and electronic maintenance/error records
  • Multiple communication interfaces: CAN, GPIB, RS232C, APG-remote
  • Leak detection and safe leak handling with low-voltage maintenance areas
  • Error detection and display system
  • Supports 1–99 replicate injections per vial
  • Low carryover: typically <0.1%, <0.05% with external needle cleaning
  • Handles sample viscosities from 0.2 to 50 centipoise
  • Fast injection cycle time: typically 50 seconds

The Agilent 1200 HPLC G1329A is designed to enhance workflow efficiency with robust injection precision, sample throughput, and integration capabilities.

Why Questpair exists

Lab equipment markets are fragmented. Buyers spend weeks calling distributors, comparing quotes, and verifying supplier reliability. Sellers struggle to reach global research labs.

We built Questpair to fix this. Every supplier on the platform is verified before listing. Every quote you receive comes from a real, vetted seller. AI prepares your case in minutes, then a real specialist takes over.

We're an independent team based in The Netherlands, working with research labs and procurement teams across Europe and worldwide.

Frequently asked questions

How long does laboratory equipment typically last?

With a service contract, ten to fifteen years is normal for most benchtop instruments. Detectors and pumps are the parts that wear first.

Can I get service for older equipment?

Often, yes. Verified suppliers frequently work with service partners converting the instrument. If applicable, we can list servicing options alongside the offer.

Where can I buy parts?

Ask a scientist through the request button. Parts are often available even when the instrument itself is discontinued.

What is the typical lead time for refurbished equipment?

Two to six weeks, depending on refurbishment scope and whether calibration is included.

Does laboratory equipment ship internationally?

We ship to 38 countries. Crating and customs handling are quoted with the instrument.

Reply in one hour

Can't find the exact configuration?

Tell us what you need and a scientist finds it for you, usually within the hour.

Last updated 25 November 2026