53k+ listingsShips to NLResponse within one hour+31 73 711 4717info@questpair.com

Agilent 1100 G1313A HPLC Autosampler

From€2,900excl. VAT

In stock

Can't find what you're looking for?

Sample Capacity
100 positions
Sample Format
Vials
Volume Range
0.1 to 100 µL
All 5 specifications
  • Ships to NL
  • Response within 1 hour
  • Warranty available. Buy without worry
  • Support from a scientist

Specifications 5

Brands
Agilent
New/Used
Used
Sample Capacity
100 positions
Sample Format
Vials
Volume Range
0.1 to 100 µL

Product description

Agilent 1100 G1313A autosampler for analytical HPLC, with 0.1 to 100 µL injections, 100 vial capacity and internal testing.

The Agilent 1100 G1313A autosampler is an analytical HPLC autosampler for pharmaceutical quality control, environmental analysis, food analysis and routine chromatography.

The autosampler handles 100 2 mL vials in one tray. It also supports 40 2 mL vials in a half tray or 15 6 mL vials in a half tray using Agilent vials. Communication interfaces include CAN, GPIB, RS232C and APG remote. Optional connections include four external contact closures and BCD vial number output.

The used autosampler was internally tested and maintained. It was refurbished if needed.

Agilent 1100 G1313A Autosampler

The Agilent 1100 G1313A is an autosampler for analytical HPLC and routine chromatography. It supports pharmaceutical quality control, environmental analysis and food analysis workflows.

Sample Handling

  • Injection range from 0.1 to 100 µL in 0.1 µL increments
  • Up to 1500 µL with multiple draw when the required hardware modification is installed
  • 1 to 99 replicate injections per vial
  • Minimum sample volume of 1 µL from 5 µL of sample in a 100 µL microvial
  • Minimum sample volume of 1 µL from 10 µL of sample in a 300 µL microvial
  • Sample viscosity range from 0.2 to 50 cP
  • Typical injection cycle time of 50 seconds, depending on draw speed and injection volume
  • Typical carryover below 0.1%
  • Typical carryover below 0.05% with external needle cleaning

Sample Capacity

  • 100 × 2 mL vials in one tray
  • 40 × 2 mL vials in a half tray
  • 15 × 6 mL vials in a half tray using Agilent vials

Precision and Control

  • Typical precision below 0.5% RSD of peak areas from 5 to 100 µL
  • Typical precision below 1% RSD of peak areas from 1 to 5 µL
  • Early maintenance feedback
  • Electronic maintenance and error records
  • Leak detection and safe leak handling
  • Low voltages in maintenance areas
  • Error detection and display

Communications

  • Controller Area Network
  • GPIB, IEEE-448
  • RS232C
  • APG remote standard
  • Optional four external contact closures
  • BCD vial number output

This used Agilent 1100 G1313A autosampler was internally tested and maintained. It was refurbished if needed.

What these specifications mean

The 0.1 to 100 µL injection range covers standard analytical HPLC sample volumes. Multiple draw can extend injection volume to 1500 µL when the required hardware modification is present.

The autosampler supports 1 to 99 replicate injections from one vial. This allows repeated injections without manually reloading the sample between runs. Typical precision is below 0.5% RSD of peak areas from 5 to 100 µL and below 1% RSD from 1 to 5 µL. Typical carryover is below 0.1%, or below 0.05% with external needle cleaning.

Why Questpair exists

Lab equipment markets are fragmented. Buyers spend weeks calling distributors, comparing quotes, and verifying supplier reliability. Sellers struggle to reach global research labs.

We built Questpair to fix this. Every supplier on the platform is verified before listing. Every quote you receive comes from a real, vetted seller. AI prepares your case in minutes, then a real specialist takes over.

We're an independent team based in The Netherlands, working with research labs and procurement teams across Europe and worldwide.

Frequently asked questions

How long does laboratory equipment typically last?

With a service contract, ten to fifteen years is normal for most benchtop instruments. Detectors and pumps are the parts that wear first.

Can I get service for older equipment?

Often, yes. Verified suppliers frequently work with service partners converting the instrument. If applicable, we can list servicing options alongside the offer.

Where can I buy parts?

Ask a scientist through the request button. Parts are often available even when the instrument itself is discontinued.

What is the typical lead time for refurbished equipment?

Two to six weeks, depending on refurbishment scope and whether calibration is included.

Does laboratory equipment ship internationally?

We ship to 38 countries. Crating and customs handling are quoted with the instrument.

Reply in one hour

Can't find the exact configuration?

Tell us what you need and a scientist finds it for you, usually within the hour.

Last updated 4 November 2026