Wavelength range
190–700 nm in steps of 1 nm, 3 wavelengths simultaneously
Bandwidth
4 nm
Wavelength accuracy
±2 nm
Wavelength reproducibility
< 500 ms (one cycle from 214 nm to 254 nm and back to 214 nm)
Linearity
< 2% deviation up to 2 AU at 260 nm with Uracil at pH 2
Availability
Sold
All 31 specifications
  • Ships to NL
  • Response within 1 hour
  • Warranty available. Buy without worry
  • Support from a scientist

Specifications 31

New/Used
Used
Wavelength range
190–700 nm in steps of 1 nm, 3 wavelengths simultaneously
Bandwidth
4 nm
Wavelength accuracy
±2 nm
Wavelength reproducibility
< 500 ms (one cycle from 214 nm to 254 nm and back to 214 nm)
Linearity
< 2% deviation up to 2 AU at 260 nm with Uracil at pH 2
Static noise
short term: < 6x10-5 AU at 230 nm long term: < 6x10-5 AU at 230 nm
Drift
< 2x10-4 AU/h at 254 nm
Optical path length 2 mm cell
2 mm, 2 µl cell volume
Optical path length 10 mm cell
10 mm, 8 µl cell volume
Max. flow rate
100 ml/min
Max. pressure
2 MPa (290 psi)
Temperature range
+4 to +40 °C (liquid)
Wetted material
PTFE (polytetrafluoroethylene) PEEK (polyetheretherketone) titanium (palladium alloy), quartz (synthetic fused silica) and titanium
Flow Rate Range
Isocratic mode: 0.01 to 100 ml/min in steps of 10 ul/min Gradient mode: 0.01 to 100 ml/min in steps of 10 ul/min Double mode: 0.01 to 200 ml/min in steps of 10 ul/min
Pressure Range
0 to 10 MPa (100 bar, 1450 psi)
pH Stability Range
1 to 13 (1 to 14 with <1 day exposure)
Viscosity
Max 5 cP
Flow Rate Accuracy
± 2% or 20 ul/min, whichever is greater
Flow Rate Reproducibility
Flow rate ≥ 0.5 ml/min: rsd < 0.5%
Gradient Composition
Accuracy: < ±1% at 0.05 to 10 ml/min Reproducibility: rsd < 0.25% at 0.5 to 100 ml/min
Pressure Sensor
Range: 0 to 27.5 MPa Offset error: < 0.05 MPa Scale error: < ±2%
Max Flow Rate
0.001 to 100 ml/min
Max Pressure
12ml chamber: 10 MPa at 0.1 to 100 ml/min (100 bar, 1450 psi) 0.6, 2, and 5ml chambers: 25 MPa at 0.001 to 100 ml/min (250 bar, 3625 psi) 90 ul and 0.2 ml chambers: 35 MPa at 0.001 to 5 ml/min (250 bar, 5075 psi)
Internal Volume
6 interchangeable mixing chambers: 90 ul, 0.2, 0.6, 2, 5, and 12 ml
Mixing Principle
1 static chamber and 1 dynamic chamber
Inlet and Outlet Tubing
UNF 10 to 32 thread profile
Fraction Size
0.1 to 99999.99 ml (volume mode) 0.1 to 99999.99 min (time mode)
Standard Mode
u-liter plates, DropSync: 0 to 1 ml/min 12 mm rack, DropSync: 0 to 1.5 ml/min 18 and 30 mm racks, DropSync: 0 to 2 ml/min 18 and 30 mm racks, Accumulator: 15 to 100ml/min
Prep Mode
30 mm rack, Accumulator: 15 to 100 ml/min Funnel-to-flask, Accumulator: 15 to 100 ml/min 250ml bottles, Accumulator: 15 to 100 ml/min
Environmental Conditions
+4° to +40°C 20 to 95% relative humidity 84 to 106 kPa (840 to 1060 mbar) atmospheric pressure

Product description

Versatile FPLC system with UV-900 detector, P-900 pump, M-925 mixer, and Frac-950 fraction collector.

The GE ÄKTA FPLC Explorer 100 System is designed for precise protein purification using:

  • Wavelength Range: 190–700 nm with 1 nm step increments
  • Wavelength Accuracy: ±2 nm
  • Max Flow Rate: 100 ml/min
  • UV-900 Detector: Simultaneous detection at three wavelengths, 4 nm bandwidth, flow cell max pressure 2 MPa
  • P-900 Pump: Accurate flow rates from 0.01 to 100 ml/min, supports isocratic and gradient modes
  • M-925 Mixer: Six interchangeable chambers, flow rate range 0.001 to 100 ml/min, max pressure up to 35 MPa
  • Frac-950 Fraction Collector: Wide range of fraction sizes, pH stability 1 to 13
  • System supports size exclusion, ion exchange, and affinity chromatography applications
  • Compatible with Unicorn 5.31 software for operation and data handling
  • Tested for full functionality including pressure sensors, valves, UV monitor, pH and conductivity measurements

This system is suitable for proteomics, drug discovery, biochemistry, molecular biology, and other protein research fields.

Why Questpair exists

Lab equipment markets are fragmented. Buyers spend weeks calling distributors, comparing quotes, and verifying supplier reliability. Sellers struggle to reach global research labs.

We built Questpair to fix this. Every supplier on the platform is verified before listing. Every quote you receive comes from a real, vetted seller. AI prepares your case in minutes, then a real specialist takes over.

We're an independent team based in The Netherlands, working with research labs and procurement teams across Europe and worldwide.

Frequently asked questions

How long does laboratory equipment typically last?

With a service contract, ten to fifteen years is normal for most benchtop instruments. Detectors and pumps are the parts that wear first.

Can I get service for older equipment?

Often, yes. Verified suppliers frequently work with service partners converting the instrument. If applicable, we can list servicing options alongside the offer.

Where can I buy parts?

Ask a scientist through the request button. Parts are often available even when the instrument itself is discontinued.

What is the typical lead time for refurbished equipment?

Two to six weeks, depending on refurbishment scope and whether calibration is included.

Does laboratory equipment ship internationally?