Easily list your scientific equipment
Top suppliers for laboratory & industry
New and used instruments
Beckman Coulter MoFlo XDP Cell Sorter Laser Cytomation System + Optics Table Lab
This system was removed from a university where it was surplus to requirement.
It is in good cosmetic condition, the working condition of each unit that we were able to test is listed below.
Parts of the system:-
Optics Table - MoFlo Optical BenchSN - 2086Dimensions - L153 W122 H80.5cmMeasurement - Imperial
Parts fitted on table:-
Pressure Console UnitSN - 2086
Coherent SapphireModel - Sapphire LP USB CDRH Driver Unit w/displayPN - 1189089SN - LDP.1189089.C02070- Powers on
Coherrent Sapphire 488-200 CW CDRHPN - 1137970SN - LDP.1137970.521683Heatsink Sapphire LP RoHS compl- NOT TESTED
Coherrent CubePN - 1170509SN - C20010803- NOT TESTED
Coherrent CubePN - 1193821/AASN - C14070901- NOT TESTED
Hamamatsu Parts:-H957-27 (No.T0050)H957-18 (No.K0149)H957-27 (No.T3721)H957-25 (No.S0319)H957-25 (No.S0360)H957-25 (No.S0359)H957-25 (No.S0322)H957-25 (No.S0324)H957-27 (No.T0051)H957-27 (No.T0054)H957-25 (No.S0358)H957-25 (No.S0321)H957-25 (No.S0323)
Smart Sampler
2 x Preamplifier UnitsSN - 2086
DTx Data Display 19" LCD w/TouchscreenBCI PN - B44222DTx PN - 90819SN - BCM0514F37003- Powers on via Miami MoFlo XDP Controller
Under table units:-
DTx BCI Miami MoFlo XDP ControllerDTx PN - 90692BCI PN - B20566DTx SN - BCM0915F44602- Powers on - When powered up, an error message reads 'OdinInstControl::connectMachine failed: AttachedInstrument::GetInstance:Repository had communication error!'
Gast Vacuum PumpDOA-P501-BN- Powers on
Intellipower UPSModel - IQ200V-C-DiAC-Isol-PFCSN - 2FKVDIFC5B15110524- Powers on
Beckman Coulter Aerosol Evacuation System Buffalo Filter w/ PedalModel - 200-0179SN - 502135Has ViroSafe Filter fitted (PN-200-8166)- Powers on.
3 x 316L UM Alloy Products Corp Pressure Vessels:-SN - S533735-1-005 (2014)SN - S514233-1-011 (2010)SN - S507165-1-008 (2008)
Other units:-
Thermo Scientific Recirculating Water BathControl - C10 Type - 003-5009/SN - 1 08004133017Bath - K20 Type - 002-4276/SN - 08004052 003- Powers on
Testing:-37˚c - Thermometer read 37.6˚c100˚c - Thermometer read 100.0˚c-20˚c - Thermometer didn't read lower than -13.5˚c
Jun Air CompressorType - OF302-25MQSN - 0119802962- Powers on with compressor running
Hours - 00898.3
Other spare bits:-1 x Black Case of various filters1 x Black Case containing 3 x Merlin Gerin fuses, Methanol & Acetone2 x Sets of Spare Shaped Charged Plates (one pair boxed)1 x Box of Filters1 x Box of mixed size O-ringsRed Ethernet cable5 x IEC Extension cables
QuestPair is a platform to find and acquire laboratory equipment by connecting research organisations, universities, and companies with high-quality, affordable instruments. The platform facilitates multiple sales daily worldwide and works with partners around the globe to provide trusted, vetted products for a wide variety of scientific needs. Our in-house developed AI algorithms collect data on the usage and availability of scientific instruments, enabling us to provide tailor-made assistance in sourcing suitable laboratory equipment. With a focus on cost-effectiveness and sustainability, QuestPair has an excellent track record in streamlining acquisitions for laboratories to further scientific research and development.