Easily list your scientific equipment
Top suppliers for laboratory & industry
New and used instruments
Memmert IF110 Forced Circulation Incubator 108L, 115V
Working Temperature Range
Min. 10°C above ambient up to +80°C
Convection Type
Forced Air
Chamber Volume
108 L (3.8 cu.ft.)
Features
- Setting temperature range from +20°C to +80°C with a working range starting at Min. 10°C above ambient
- Volume of 108 L (3.8 cu.ft.) with a reinforced stainless steel interior
- ControlCOCKPIT adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-color display
- Adjustable parameters include temperature (Celsius or Fahrenheit), fan speed, air flap position, and program time
- Forced air circulation by quiet air turbine, adjustable in 10% steps with pre-heated fresh air admixture
- Safety features include an adjustable overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880
- Easy-to-clean stainless steel interior reinforced with deep drawn ribbing and integrated large-area heating on four sides
The Memmert IF110 Forced Circulation Incubator 108L, 115V is designed to provide precise temperature control and reliable performance for laboratory applications. With a setting temperature range from +20°C to +80°C and a working range starting at 10°C above ambient, this incubator ensures stable thermal conditions for a wide variety of processes. Its spacious 108L (3.8 cu.ft.) chamber is constructed with reinforced stainless steel, offering both durability and easy cleaning. The advanced ControlCOCKPIT adaptive multifunctional digital PID-microprocessor controller, featuring a high-definition TFT-color display, allows for easy adjustments to critical parameters like temperature, fan speed, air flap position, and program time.
The IF110 also prioritizes safety with an adjustable overtemperature monitor and mechanical temperature limiter (TB) that meets protection class 1 standards according to DIN 12880. Forced air circulation, powered by a quiet air turbine adjustable in 10% steps, ensures uniform temperature distribution, while the pre-heated fresh air admixture feature allows for greater environmental control. Additional functionality includes the AtmoCONTROL software for programming and data management via Ethernet or USB. With a maximum chamber load of 385 lbs and each shelf supporting up to 44 lbs, the IF110 incubator is a reliable and efficient solution for demanding laboratory environments.
INCLUDED:
TheMemmert IF110 Forced Circulation Incubator 108L, 115V includes:
- 2 Stainless Steel Grids
- Tilt Protection Wall Bracket
- Calibration Certificate
- Power Cord
- Operator's Manual
- Manufacturer's Documentation
- 3 Year Manufacturer's Warranty
Setting Temperature Range | +20 to +80°C |
Working Temperature Range | Min. 10°C above ambient up to +80°C |
Setting Accuracy (temperature) | 0.1°C |
Volume | 108 L (3.8 cu.ft.) |
Temperature Sensor | 1 Pt100 sensor DIN class A in 4-wire-circuit |
Control Panel | ControlCOCKPIT: Adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-color display |
Timer | Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
Adjustable Parameters | Temperature (Celsius or Fahrenheit), fan speed, air flap position, program time, time zones, daylight savings time |
Function SetpointWAIT | The process time does not start until the set temperature is reached |
Language Setting | German/English/Spanish/French |
Calibration | Three freely selectable temperature values |
Fan | Forced air circulation by quiet air turbine, adjustable in 10% steps |
Fresh Air Admixture | Adjustment of pre-heated fresh air admixture by air flap control in 10% steps |
Vent | Vent connection with restrictor flap |
Programming | AtmoCONTROL software for managing and organizing data via Ethernet, USB stick available as accessory |
Documentation | Program stored in case of power failure |
Temperature Control | Adjustable electronic overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880 |
Autodiagnostic System | For fault analysis |
Door | Fully insulated stainless steel door with 2-point locking (compression door lock), inner glass door |
Interior | Easy-to-clean stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides |
Max. Number of Shelves | 5 |
Max. Loading Per Shelf | 44 lbs |
Max. Loading of Chamber | 385 lbs |
Interior Dimensions (WxHxD) | 22” x 18.9” x 15.8” (56 x 48 x 40 cm) |
Exterior Dimensions (WxHxD) | 29.3” x 34” x 23” (74.5 x 86.4 x 58.4 cm) |
Net Weight | Approx. 168 lbs |
Electrical Data | 115 V, 50/60 Hz, approx. 900 W |
QuestPair is a platform to find and acquire laboratory equipment by connecting research organisations, universities, and companies with high-quality, affordable instruments. The platform facilitates multiple sales daily worldwide and works with partners around the globe to provide trusted, vetted products for a wide variety of scientific needs. Our in-house developed AI algorithms collect data on the usage and availability of scientific instruments, enabling us to provide tailor-made assistance in sourcing suitable laboratory equipment. With a focus on cost-effectiveness and sustainability, QuestPair has an excellent track record in streamlining acquisitions for laboratories to further scientific research and development.
Customer Service
Here to help you with: Product Inquiries, Shipping & Support, Technical Support, Business Inquiries and Press.
We are available to assist you Mon-Fri, 10am - 5pm CET.