The FACSCanto II system is an easy-to-use benchtop analyzer that delivers proven performance, accuracy, and high-quality results. The FACSCanto II can be configured with two or three lasers to detect up to eight colors. It features many innovations, including a true fixed alignment flow cell to minimize startup time and improve reproducibility. The optical system maximizes signal detection and increases sensitivity and resolution for each color in a multicolor assay. These and other capabilities make the BD FACSCanto II ideal for today’s busy clinical lab, providing a high degree of automation and quality control. With optimal reproducibility, the FACSCanto II system reduces hands-on technician time and costs associated with repeat testing.
Lasers:
405-nm solid state diode, 30-mW fiber power output
488-nm solid state, 20-mW laser output
633-nm HeNe, 17-mW laser output
Laser Configuration: Spatially separated beams with 9 x 65-
μm elliptical spots
Optical Alignment Procedure: Fixed, no operator alignment required
Flow Cell: 180 x 430-μm rectangular quartz flow cell
Collection Optics: Optical-gel coupled 1.2 NA lens
FSC Resolution: 1.0 μm
SSC Resolution: 0.5 μm
Fluorescence Detector Design: Reflective optics with single transmission filter in front of each PMT
FSC Detector: Photodiode with 488/10 BP
SSC Detector: PMT with 488/10 BP
Fluorescence Detectors: 8 PMTs in 4-2-2 configuration
Violet Laser Dyes: Pacific Blue, AmCyan (455, 488 nm)
Blue Laser Dyes: FITC, PE, PerCP or PerCP-Cy 5.5, PE-Cy 7 (525, 575, 678 or 695, 785 nm)
Red Laser Dyes: APC, APC-Cy7 (660, 785 nm)
Detector Bands:
Violet Laser: 450/50; 502 to 525 nm
Blue Laser: 530/30; 585/42; >670; 780/60 nm
Red Laser: 660/20; 780/60 nm
Fluorescence Threshold Sensitivities: FITC <100 MESF; PE <50 MESF
Max Particle Size: 50 μm
Sample Flow Rate: 10 μL/min to 120 μL/min
Sample Acquisition Rate: 10,000 events/second, 6 compensated
fluorescence parameters and 2 scatter parameters
Sample Dead Volume: 30 μL (BD Falcon tubes 12 x 75-mm)
Dimensions (DxWxH):
Cytometer: 610 x 910 x 640 mm
Fluidics cart: 610 x 790 x 640 mm
Weight:
Cytometer: 145 kg
Fluidics cart: 51 kg
QuestPair is a platform to find and acquire laboratory equipment by connecting research organisations, universities, and companies with high-quality, affordable instruments. The platform facilitates multiple sales daily worldwide and works with partners around the globe to provide trusted, vetted products for a wide variety of scientific needs. Our in-house developed AI algorithms collect data on the usage and availability of scientific instruments, enabling us to provide tailor-made assistance in sourcing suitable laboratory equipment. With a focus on cost-effectiveness and sustainability, QuestPair has an excellent track record in streamlining acquisitions for laboratories to further scientific research and development.