System contains:
Solvent tray
Vacuum degasser: G1322A
Vacuum degasser:
The Agilent 1100 HPLC Degasser features a 4-channel vacuum container, including four tubular plastic membranes, and a vacuum pump. This degasser has a maximum flow rate of 0-10ml/min flow rate per channel and 5-10ml/min at reduced degassing performance.
Quaternary pump:
Hydraulic system: Dual plunger in series pump with proprietary servo-controlled variable stroke drive, floating plungers and active inlet valve
Setable flow range: 0.001 – 10 ml/min, in 0.001 ml/min increments
Flow range: 0.2 – 10.0 ml/min
Flow precision: < 0.07% RSD, or < 0.02 min SD whatever is greater, based on retention time at constant room temperature
Flow accuracy: ± 1% or 10 µl/min whatever is greater
Pressure: Operating range 0 – 400 bar up to 5 ml/min, 0 – 200 bar up to 10 ml/min
Pressure pulsation: < 2 % amplitude (typically < 1 %), at 1 ml/min isopropanol, at all pressures > 10 bar
Compressibility compensation: User-selectable, based on mobile phase compressibility
Recommended pH range: 1.0 – 12.5, solvents with pH < 2.3 should not contain acids which attack stainless steel
Gradient formation: Low pressure quaternary mixing/gradient capability using proprietary high-speed proportioning valve Delay volume 800 – 1100 µl, dependent on back pressure
Composition Range: 0 – 95 % or 5 – 100 %, user selectable
Composition Precision: < 0.2 % RSD, at 0.2 and 1 ml/min
Autosampler thermostat:
Temperature range: setable from 4°C to 40°C in 1°C increments
Temperature accuracy (at ambient temperatures < 25 °C and humidity < 50%): – 1°C to + 4°C at a setpoint of 4°C
Temperature accuracy (at ambient temperatures > 25 °C and/or humidity > 50%): – 1°C to + 5°C at a setpoint of 4°C
Standard autosampler:
Pressure: Operating range 0 – 400 bar
GLP features: Early maintenance feedback (EMF), electronic records of maintenance and errors
Safety features: Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display
Injection range: 0.1 – 100 µl in 0.1 µl increments
Replicate injections: 1 – 99 from one vial
Precision: <0.25% RSD from 5 – 100 µl, <1% RSD 1 – 5 µl
Minimum sample volume: 1 µl from 5 µl sample in 100 µl microvial, or 1 µl from 10 µl sample in 300 µl microvial
Carryover: Typically <0.1%
Sample viscosity range: 0.2 – 50 cp
Replicate injections per vial: 1 – 99
Sample capacity 100 x 2 ml vials in 1 tray
Injection cycle time: Typically 50 s depending on draw speed and injection volume
Thermostatted colomn compartment:
Temperature range: 10 degrees below ambient to 80 °C, up to 80 °C: flow rates up to 5 ml/min
Temperature stability: ± 0.15 °C
Temperature accuracy: ± 0.8 °C (± 0.5 °C With calibration)
Column capacity: Three 30 cm
Warm-up/cool-down time: 5 minutes from ambient to 40 °C, 10 minutes from 40 – 20 °C
Dead volume: 3 µl
Communications: Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN via other series module
Safety and maintenance: Extensive diagnostics, error detection and display (through control module and Agilent ChemStation), leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas.
Diode array detector:
Detection type: 1024-element photodiode array
Light source: Deuterium and tungsten lamps
Wavelength range: 190 – 950 nm
Short term noise (ASTM) Single and Multi-Wavelengt: ± 1 × 10-5 AU at 254 and 750 nm
Drift: 2 × 10-3 AU/hr at 254 nm
Linear absorbance range: > 2 AU (upper limit)
Wavelength accuracy: ± 1 nm Self-calibration with deuterium lines, verification with holmium oxide filter
Wavelength bunching: 1 – 400 nm Programmable in steps of 1 nm
Slit width: 1, 2, 4 , 8, 16 nm Programmable slit
Diode width: < 1 nm
Flow cells: 13 µl volume, 10 mm cell path length and 120 bar (1760 psi) pressure maximum
Control and data evaluation: Agilent ChemStation for LC
Analog outputs: Recorder/integrator: 100 mV or 1 V, output range 0.001 – 2 AU, two outputs
Communications: Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN optional
Safety and maintenance: Extensive diagnostics, error detection and display through control module and ChemStation), leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas.
GLP features: Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of lamp burn time with user-settable limits and feedback messages. Electronic records of maintenance and errors. Verification of wavelength accuracy with built-in holmium oxide filter.
QuestPair is a platform to find and acquire laboratory equipment by connecting research organisations, universities, and companies with high-quality, affordable instruments. The platform facilitates multiple sales daily worldwide and works with partners around the globe to provide trusted, vetted products for a wide variety of scientific needs. Our in-house developed AI algorithms collect data on the usage and availability of scientific instruments, enabling us to provide tailor-made assistance in sourcing suitable laboratory equipment. With a focus on cost-effectiveness and sustainability, QuestPair has an excellent track record in streamlining acquisitions for laboratories to further scientific research and development.